指輪型の非侵襲的連続血圧計測システム開発研究-修正規準化脈波容積を用いたボリュームクランプ法における平均動脈血圧の決定法に関する予備検討-
http://hdl.handle.net/11478/1490
http://hdl.handle.net/11478/149081884d8d-2615-4559-a310-2d14b22454d9
名前 / ファイル | ライセンス | アクション |
---|---|---|
11478-1490_p99.pdf (2.0 MB)
|
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2020-04-24 | |||||
タイトル | ||||||
タイトル | 指輪型の非侵襲的連続血圧計測システム開発研究-修正規準化脈波容積を用いたボリュームクランプ法における平均動脈血圧の決定法に関する予備検討- | |||||
タイトル | ||||||
タイトル | Development of the Ring-type Non-invasive Continuous Blood Pressure Monitoring System -A Pilot Study on the Determination Method of Mean Arterial Pressure Based on the Volume Clamp Method Using the Modified Normalized Pulse Volume- | |||||
言語 | en | |||||
言語 | ||||||
言語 | jpn | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | modified normalized pulse volume | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | pulse volume | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | mean arterial pressure | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | finger photo-photoplethysmogram | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | modified normalized pulse volume | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | pulse volume | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | mean arterial pressure | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | finger photo-photoplethysmogram | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
著者 |
李, 知炯
× 李, 知炯 |
|||||
著者(ヨミ) | ||||||
識別子Scheme | WEKO | |||||
識別子 | 2614 | |||||
姓名 | リ, ジヒョン | |||||
別言語の著者 | ||||||
識別子Scheme | WEKO | |||||
識別子 | 2615 | |||||
姓名 | LEE, Jihyoung | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | Along with the development of non-invasive techniques for the high accuracy measurement of blood pressure on the finger, the investigation of finger pulse volume (PV) variability has been attracted considerable interest. Modified normalized pulse volume (mNPV) has the potential to be a practical index for the determination method of mean arterial pressure (MAP) based on the volume clamp method due to its very sensitive to vasoconstrictor stimuli on the finger. However, MAP derived by mNPV has not yet been investigated. Therefore, we compared MAP measured by sphygmomanometer as a reference, with MAP derived by the PV and mNPV using the developed prototype system during baseline in ten young subjects. The results gave mean differences in Bland-Altman plots between the measured MAP and MAP derived by PV and mNPV of 1.86 ± 6.49 (standard deviation) and -2.54 ± 8.12, respectively. mNPV is discussed in terms of the availability for determination of MAP based on VCM. In conclusion, mNPV is a valid approach, which might be useful for the calibration index of MAP derived by PV. | |||||
書誌情報 |
福岡工業大学総合研究機構研究所所報 巻 2, p. 99-103, 発行日 2020-04-04 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 24345725 | |||||
フォーマット | ||||||
内容記述タイプ | Other | |||||
内容記述 | application/pdf | |||||
形態 | ||||||
値 | 1992477 bytes | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
タイトル(ヨミ) | ||||||
その他のタイトル | ユビワガタ ノ ヒシンシュウテキ レンゾク ケツアツ ケイソク システム カイハツ ケンキュウ - シュウセイ キジュンカ ミャクハ ヨウセキ オ モチイタ ボリューム クランプホウ ニ オケル ヘイキン ドウミャク ケツアツ ノ ケッテイホウ ニ カンスル ヨビ ケントウ - | |||||
出版者 | ||||||
出版者 | 福岡工業大学総合研究機構 | |||||
出版者(ヨミ) | ||||||
値 | フクオカ コウギョウ ダイガク ソウゴウ ケンキュウ キコウ | |||||
資源タイプ | ||||||
内容記述タイプ | Other | |||||
内容記述 | 論文(Article) | |||||
資源タイプ・ローカル | ||||||
値 | 紀要論文 | |||||
資源タイプ・NII | ||||||
値 | Departmental Bulletin Paper | |||||
資源タイプ・DCMI | ||||||
値 | text | |||||
資源タイプ・ローカル表示コード | ||||||
値 | 02 | |||||
コメント | ||||||
値 | 情報科学研究所(Computer Science Laboratory) |